Biobase Lab Air Protect Furniture Class II B2 Biological Safety Cabinet

Product Details
Customization: Available
Capacity: Two Person
After-sales Service Provided: Engineers available to service machinery overseas

360° Virtual Tour

Diamond Member Since 2019

Suppliers with verified business licenses

Audited Supplier

Audited by an independent third-party inspection agency

to see all verified strength labels (25)
  • Biobase Lab Air Protect Furniture Class II B2 Biological Safety Cabinet
  • Biobase Lab Air Protect Furniture Class II B2 Biological Safety Cabinet
  • Biobase Lab Air Protect Furniture Class II B2 Biological Safety Cabinet
  • Biobase Lab Air Protect Furniture Class II B2 Biological Safety Cabinet
  • Biobase Lab Air Protect Furniture Class II B2 Biological Safety Cabinet
  • Biobase Lab Air Protect Furniture Class II B2 Biological Safety Cabinet
Find Similar Products
  • Overview
  • Product Description
  • Product Parameters
  • Related products
  • Company Profile
  • Packaging & Shipping
  • Contact
Overview

Basic Info.

Model NO.
BSC-1300IIB2-X
Certification
/
Usage
Research Institute, Hospital, School, Company
Noise
≤62DB
Type
Biological Safety Cabinet
Internal Size (W*D*H)
1150*600*660 mm
External Size (W*D*H)
1300*750*2250mm
Tested Opening
Safety Height 200 mm
Max Opening
420mm
Inflow Velocity
0.53+-0.025 M/S
Down Flow Velocity
0.33+-0.025 M/S
Pre-Filter
Washable
HEPA Filter
Two, 99.999% Efficiency
Front Window
Motorized
UV Lamp
30W* 1
Illuminating Lamp
LED Lamp
Consumption
850W
Waterproof Sockets
2
Display
LCD Display
Control System
Microprocessor
Transport Package
Wooden Case or Carton Package
Specification
1300*750*2250mm
Trademark
BIOBASE
Origin
China
Production Capacity
1000 Units/Month

Product Description


BIOBASE Class II B2 Biological Safety Cabinet 

Product Description

 

Applications of class II B2 biological safety cabinet
Three protection: operator, sample and environment.
Airflow system: 0 % air recirculation, 100% air exhaust
B2 Biological Safety Cabinet has higher class of safety protection compared with type A2,it may be used for work with volatile toxic chemicals and radionuclides as required as adjuncts to microbiological studies.

Biobase Lab Air Protect Furniture Class II B2 Biological Safety Cabinet

 

Advantages of class II B2 biological safety cabinet

1. Motorized Front Window. 

2. Time Reserve Function.

3. HEPA Filter life and UV Life indicator. 

4. Automatic air speed adjustable with filter block.

5. Work area surrounded by negative pressure, it can ensure maximum safety in work area.

6. Audio and visual alarm (Filter replacement, Window over height, abnormal Air flow velocity etc).

7. Most accessories are standard. No need to pay more, save your money.

8. Remote Control. All functions can be realized with it, making the operation much easier and more convenient.

9. Foot Switch. Adjust front window height by foot during experiment, to avoid airflow turbulence caused by arm movement.

10. Interlock Function: UV lamp and front window; UV lamp and blower, fluorescent lamp; Blower and front window.

Product Parameters


Technical Parameters of class II B2 biological safety cabinet

Model

BSC-1100IIB2-X

BSC-

1300IIB2-X

BSC-

1500IIB2-X

BSC-

1800IIB2-X

Internal Size

(W*D*H)

940*600*660 mm

1150*600

*660 mm

1350*600

*660mm

1670*600

*660 mm

External Size

(W*D*H)

1100*750*2250mm

1300*750

*2250mm

1500*760

*2250mm

1820*750

*2250mm

Tested Opening

Safety Height = 200 mm(8'')
 

Max Opening

420mm(17'')

420mm(17'')

500mm(20'')

480mm

(20'')

Inflow Velocity

0.53±0.025 m/s

Down flow Velocity

0.33±0.025 m/s

HEPA Filter

Two, 99.999% efficient at 0.3 um. Filter life indicator.

Front Window

Motorized. Two-layer laminated toughened glass ≥ 5mm. Anti UV.

Noise

EN1246949 ≤ 58 dB / NSF49 ≤ 61 dB

UV Lamp

30W*1

30W* 1

40W* 1

40W*2

UV timer, UV life indicator, emission of 253.7 nanometers 

for most efficient decontamination.

Illuminating Lamp

LED Lamp

LED Lamp

LED Lamp

Fluorescent

Lamp

12W*2

14W*2

16W*2

40W*2

Illumination

≥1000Lux

Consumption

700W

850W

900W

1200W

Waterproof Sockets

Two, total load of two sockets: 500W

Display

LCD display: exhaust filter and down flow filter pressure,

filter and UV lamp working time, 

inflow and down flow velocity, filter life,

humidity and temperature, system working time etc.

Control System

Microprocessor

Airflow System

0% air recirculation, 100% air exhaust

Alarm

Abnormal airflow velocity; Filter replacement;

Front window at unsafe height.

Exhaust Duct

4 meters PVC duct, Diameter:300mm

Material

Work Zone: 304 stainless steel

Main Body: Cold-rolled steel with anti-bacteria powder coating.

Work Surface Height

750mm

Caster

Foot master caster

Power Supply

AC 220V±10%, 50/60Hz; 110V±10%,

60Hz (110V/60Hz is not applicable to BSC-1800IIB2-X)

Standard Accessory

Illuminating lamp, UV lamp*2,base stand, remote control,

 foot switch, exhaust blower, 

exhaust duct, drain valve, waterproof sockets*2, pipe strap*2

Optional Accessory

Water and gas tap, Electric height adjustable base stand

Gross Weight

246kg

276kg

302kg

408kg

Package

Main Body

1230* 990* 1810 mm

1460*1050

*1800 mm

1650*990

*1810 mm

1950*990*

1950 mm

Exhaust Blower (W*D*H)

970* 810* 630 mm

970* 810

* 630 mm

970* 810

* 630 mm

920* 710

* 680 mm




 

LCD Display                                                                    Work Zone

Large digital display is easy to monitor all the                Work zone, made of 304 stainless steel,

safety parameters at a glance and ergonomically           is surrounded by negative pressure.

sized control panel improves user interface.


 

Remote Control                                                             UV lamp

All functions can be realized with it, making the            Emission of 253.7 nanometers for most efficient

operation much easier and more convenient.               decontamination.
 

Front Window                                                                 Footmaster caster

Two-layer laminated toughened glass>5mm,                 Universal caster with brake and leveling feet.

anti UV                                                                            Adjust front window height by foot during

                                                                                       experiment, to avoid airflow turbulence                                                                                                                       caused by arm movement.                                                                               
 

Foot Switch                                                                      Waterproof Socket

Adjust front window height by foot during                         2 waterproof sockets are located in the 

experiment, to avoid airflow turbulence caused                side wall, for optimum convenience of

by arm movement.                                                            using small devices inside the cabinet.
 

Exhaust Blower

 

Related products


BIOBASE Products
Biobase Lab Air Protect Furniture Class II B2 Biological Safety Cabinet

Company Profile

 

BIOBASE Company
Biobase Lab Air Protect Furniture Class II B2 Biological Safety Cabinet
Biobase Lab Air Protect Furniture Class II B2 Biological Safety Cabinet

BIOBASE Showing Room 
Biobase Lab Air Protect Furniture Class II B2 Biological Safety Cabinet


BIOBASE ExhibitionsBiobase Lab Air Protect Furniture Class II B2 Biological Safety Cabinet

 

Packaging & Shipping


Biobase Lab Air Protect Furniture Class II B2 Biological Safety Cabinet

Contact

Biobase Lab Air Protect Furniture Class II B2 Biological Safety Cabinet

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now
Contact Supplier
Product Groups
More