Customization: | Available |
---|---|
Material: | Stainless Steel |
Type: | Medicine Cabinet |
Still deciding? Get samples of US$ 4380/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Application:
Three protection: operator, sample and environment.
Airflow system: 70% air recirculation, 30% air exhaust
A2 Cabinet is suitable for working with microbiological research in the absence of volatile or toxic chemicals and radionuclide.
Advantages:
1. Motorized front window.
2. Large LCD display, all information displayed.
3. Automatic air speed adjustable with filter block.
4. With memory function in case of power-failure.
5. Side & back wall is made up of single piece stainless steel.
6. Interlock function: UV lamp and front window; UV lamp and blower, fluorescent lamp; Blower and front window.
7. Front 10° slanted to offer operator comfort while working for long time, reduce glare and maximize reach into the work area.
Model |
BSC-4FA2(4') |
BSC-4FA2-GL(4') |
BSC-6FA2-GL(6') |
BSC-4FA2-NA(4') |
|
External Size(W*D*H) |
1383*775*2295 mm |
1383*775*229 5mm |
1873*775*2295 mm |
1383*775*2295 mm |
|
Internal Size(W*D*H) |
1210*660*660mm |
1210*600*660 mm |
1700*600*660mm |
1210*600*660 mm |
|
Work Surface Height |
750mm |
||||
Max Opening |
550mm (22'') |
||||
Tested Opening |
Safety height 200mm (8'') |
Safety height 254mm (10'') |
|||
Airflow Volume |
Inflow |
462m3/h (272cfm) |
587m3/h (346cfm) |
824m3/h (485cfm) |
587m3/h (346cfm) |
Down flow |
65%:863m3/h(507cfm) |
60%:863m3/h (508cfm) |
60%:1212m3/h (713cfm) |
60%:863m3/h (507cfm) |
|
Exhaust |
35%:462m3/h(272cfm) |
40%:587m3/h (346cfm) |
40%:824m3/h (485cfm) |
40%:587m3/h (346cfm) |
|
Front Window |
Motorized. Two-layer laminated toughened glass ≥5mm. Anti UV. |
||||
HEPA Filter |
Two, 99.999% efficiency at 0.3um, filter life indicator |
||||
Filter Guard Type |
Aluminum alloy frame |
||||
Noise |
NSF 49≤61 dB/EN 12469≤58 dB |
||||
Illumination |
≥1000Lux |
||||
Display |
LCD Display |
||||
Waterproof Socket |
Two, Total consumption:≤500W |
||||
Tap |
Water tap*1, Gas tap *1 |
||||
Ground Resistance |
≤ 0.10Ω |
||||
Power Supply |
AC220V±10%, 50/60Hz; 110V±10%, 60Hz, Full load Amps: 9A, BTU/Hr: 1689 |
||||
Consumption |
400W |
400W |
500W |
400W |
|
Alarm |
Abnormal airflow velocity; Filter replacement; Front window at unsafe height. |
||||
DC Motor |
One ECM motor, 110V& 220V acceptable, Speed adjustable, high efficiency and low power consumption |
||||
Material |
Working area: 304stainless steel Frame and Decorative Plate: cold-rolled steel with anti-bacteria power coating. |
||||
UV Lamp |
30W |
30W |
40W |
30W |
|
Germicidal UV lamp, Emission of 253.7 nanometers for most efficient decontamination |
1.Why do we cooperate with BIOBASE? |
BIOBASE group, founded in 1999 and been through high and stable development for 15 years.The products we manufacture extends to most of the laboratory items and scientific analytical equipments. Our mission is to enable our customers to make the world healthier, cleaner and safer. |
2.How about BIOBASE payment term? |
T/T in advance & L/C & (70% deposit,the balance before shipment). |
3.How about BIOBASE delivery time? |
10-15 days after we receive prepayment. |
4.How about the package? |
Plywood case or carton or standard export wooden case. |
5.What's the loading port? |
FOB to Qingdao port China or as customer's request |