Customization: | Available |
---|---|
Classification: | Biochemical Apparatus |
Type: | Biochemistry Analyzer |
Still deciding? Get samples of US$ 3000/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Model |
BK-200 |
|
Overall |
Throughput |
200 tests/hour |
Analysis Method |
End-point, Fixed-time, Rate(Kinetic), Turbidimetry |
|
Certificates |
|
|
Sample&Reagent System |
Sample Positions |
37 sample positions |
Reagent Position |
28 reagent positions |
|
Sample Volume |
2~70µl |
|
Reagent Volume |
20~350µl |
|
Probe |
with anti-collision function, liquid level detection function |
|
Probe Washing |
Automatic washing interior and exterior |
|
Reagent Cooling |
Refrigerated tray with independent switch |
|
Reaction System |
Temperature Control |
37±0.2ºC |
Cuvettes |
48 reusable cuvettes, optical length 6mm |
|
Mixer Probe |
Independent stirring |
|
Washing |
Automatic cuvettes washing |
|
STAT Function |
YES |
|
Optical System |
Light Source |
6V/10W halogen lamp |
Spectrophotometry |
Post-spectral spectrophotometry |
|
Wavelength |
340, 405, 450, 510, 546, 578, 630, 700nm |
|
Absorbance |
0~3.0Abs |
|
Calibration&QC |
Calibration |
Linear: K factor, 1-point, 2-point and multipoint linear Non-Linear: Spline, Polygon, Logit-4P, Logit-5P |
Quality Control |
Real-time QC, Westgard multi rule, Cumulative sum check, Twin Plot(2D), Daily QC, Monthly QC |
|
Data Management |
Software |
Windows 7/8/10, 32 or 64 bit |
LIS System |
Bi-direction, support HL7 protocol |
|
Interface |
LAN port access |
|
Printer |
External optional, multiple reporting mode available |
|
Working Conditions |
Power Supply |
AC220V±10%, 50/60Hz; 110V±10%, 60Hz, 300W |
Temperature |
15~30ºC |
|
Water Consumption |
Deionized water:<5L/H |
|
Humidity |
40%~85% |
|
Size&Weight |
External Size (W*D*H) |
625*425*460mm |
Net Weight |
36kg |
|
Package Size (W*D*H) |
750*560*900mm |
|
Gross Weight |
55kg |