Customization: | Available |
---|---|
After-sales Service: | 1year |
Warranty: | 1 Year |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
MicroPette Mechanical Pipette
All pipettes have been quality tested according to ISO8655-2:2002 with calibration certificate. The quality control involves gravimetric testing of each pipette with distilled water at 22ºC.
Fully Autoclave:
The MicroPette plus pipettes can be fully autoclaved at 121ºC, 1 atm for minutes. The pipettes can be autoclaved without special preparations. After autoclaving the pipette must be cooled down and dried over 12 hours.
Features for TopPette
1. Light weight, ergonomic, low force design
2. Digital display clearly reads volume setting
3. The pipettes cover volume range of 0.1μl to 10ml
4. Easy to calibrate and maintain
5. The lower part is autoclavable
6. Design helps avoid repetitive strain injuries Features for Multi-channel pipette
7. 8 and 12 channel pipettes are available for standard 96-well plate
8. Dispensing head rotates ensure optimum pipetting convenience
9. Individual piston/tip cone assemblies allow easy repair and maintenance
10. Compound material tip cone design allows visual seal vertification
11. Universal pipette tips can be used
Single-channel Adjustable Volume Pipettes |
||||||
Volume Range |
Increment |
Test Volume (μl) |
Maximume Permissible Systematic error (Inaccuracy) |
Maximume Permissible Random error (Inaccuracy) |
||
% |
μl |
% |
μl |
|||
0.1-2.5μl |
0.05μl |
2.5 |
2.50% |
0.0625 |
2.00% |
0.05 |
1.25 |
3.00% |
0.0375 |
3.00% |
0.0375 |
||
0.25 |
12.00% |
0.03 |
6.00% |
0.015 |
||
0.5-10μl |
0.1μl |
10 |
1.00% |
0.1 |
0.80% |
0.08 |
5 |
1.50% |
0.075 |
1.50% |
0.075 |
||
1 |
2.50% |
0.025 |
1.50% |
0.015 |
||
2-20μl |
0.5μl |
20 |
0.90% |
0.18 |
0.40% |
0.08 |
10 |
1.20% |
0.12 |
1.00% |
0.1 |
||
2 |
3.00% |
0.06 |
2.00% |
0.04 |
||
5-50μl |
0.5μl |
50 |
0.60% |
0.3 |
0.30% |
0.15 |
25 |
0.90% |
0.225 |
0.60% |
0.15 |
||
5 |
2.00% |
0.1 |
2.00% |
0.1 |
||
10-100μl |
1μl |
100 |
0.80% |
0.8 |
0.15% |
0.15 |
50 |
1.00% |
0.5 |
0.40% |
0.2 |
||
10 |
3.00% |
0.3 |
1.50% |
0.15 |
||
20-200μl |
1μl |
200 |
0.60% |
1.2 |
0.15% |
0.3 |
100 |
0.80% |
0.8 |
0.30% |
0.3 |
||
20 |
3.00% |
0.6 |
1.00% |
0.2 |
||
50-200μl |
1μl |
200 |
0.60% |
1.2 |
0.15% |
0.3 |
100 |
0.80% |
0.8 |
0.30% |
0.3 |
||
50 |
1.00% |
0.5 |
0.40% |
0.2 |
||
100-1000μl |
5μl |
1000 |
0.60% |
6 |
0.20% |
2 |
500 |
0.70% |
3.5 |
0.25% |
1.25 |
||
100 |
2.00% |
2 |
0.70% |
0.7 |
||
200-1000μl |
5μl |
1000 |
0.60% |
6 |
0.20% |
2 |
500 |
0.70% |
3.5 |
0.25% |
1.25 |
||
200 |
0.90% |
1.8 |
0.30% |
0.6 |
||
1000-5000μl |
50μl |
5000 |
0.50% |
25 |
0.15% |
7.5 |
2500 |
0.60% |
15 |
0.30% |
7.5 |
||
1000 |
0.70% |
7 |
0.30% |
3 |
||
2-10ml |
0.1ml |
10ml |
0.60% |
60 |
0.20% |
20 |
5ml |
1.20% |
60 |
0.30% |
15 |
||
2ml |
3.00% |
60 |
0.60% |
12 |
8-channel Adjustable Volume Pipettes |
||||||||
Volume Range |
Increment |
Test Volume (μl) |
Maximume Permissible Systematic error (Inaccuracy) |
Maximume Permissible Random error (Inaccuracy) |
||||
% |
μl |
% |
μl |
|||||
0.5-10μl |
0.1μl |
10 |
1.50% |
0.15 |
1.50% |
0.15 |
||
5 |
2.50% |
0.125 |
2.50% |
0.125 |
||||
1 |
4.00% |
0.04 |
4.00% |
0.04 |
||||
5-50μl |
0.5μl |
50 |
1.00% |
0.5 |
0.50% |
0.25 |
||
25 |
1.50% |
0.375 |
1.00% |
0.25 |
||||
5 |
3.00% |
0.15 |
2.00% |
0.1 |
||||
50-300μl |
5μl |
300 |
0.70% |
2.1 |
0.25% |
0.75 |
||
150 |
1.00% |
1.5 |
0.50% |
0.75 |
||||
50 |
1.50% |
0.75 |
0.80% |
0.4 |
12-channel Adjustable Volume Pipettes |
||||||||
Volume Range |
Increment |
Test Volume (μl) |
Maximume Permissible Systematic error (Inaccuracy) |
Maximume Permissible Random error (Inaccuracy) |
||||
% |
μl |
% |
μl |
|||||
0.5-10μl |
0.1μl |
10 |
1.50% |
0.15 |
1.50% |
0.15 |
||
5 |
2.50% |
0.125 |
2.50% |
0.125 |
||||
1 |
4.00% |
0.04 |
4.00% |
0.04 |
||||
5-50μl |
0.5μl |
50 |
1.00% |
0.5 |
0.50% |
0.25 |
||
25 |
1.50% |
0.375 |
1.00% |
0.25 |
||||
5 |
3.00% |
0.15 |
2.00% |
0.1 |
||||
50-300μl |
5μl |
300 |
0.70% |
2.1 |
0.25% |
0.75 |
||
150 |
1.00% |
1.5 |
0.50% |
0.75 |
||||
50 |
1.50% |
0.75 |
0.80% |
0.4 |
Fixed Volume Pipettes |
||||||||
Volume Range |
Increment |
Test Volume (μl) |
Maximume Permissible Systematic error (Inaccuracy) |
Maximume Permissible Random error (Inaccuracy) |
||||
% |
μl |
% |
μl |
|||||
5μl |
- |
5μl |
1.3% |
0.065 |
1.2% |
0.06 |
||
10μl |
- |
10μl |
0.8% |
0.08 |
0.8% |
0.08 |
||
20μl |
- |
20μl |
0.6% |
0.12 |
0.5% |
0.1 |
||
25μl |
- |
25μl |
0.5% |
0.125 |
0.3% |
0.075 |
||
50μl |
- |
50μl |
0.5% |
0.25 |
0.3% |
0.15 |
||
100μl |
- |
100μl |
0.5% |
0.5 |
0.3% |
0.3 |
||
200μl |
- |
200μl |
0.4% |
0.8 |
0.2% |
0.4 |
||
250μl |
-- |
250μl |
0.4% |
1.0 |
0.2% |
0.5 |
||
500μl |
- |
500μl |
0.3% |
1.5 |
0.2% |
1.0 |
||
1000μl |
- |
1000μl |
0.3% |
3.0 |
0.2% |
2.0 |
||
2000μl |
- |
2000μl |
0.3% |
6.0 |
0.15% |
3.0 |
||
5000μl |
- |
5000μl |
0.3% |
15 |
0.15% |
7.5 |
Company Profile
Packaging & Shipping
Exhibition