Customization: | Available |
---|---|
Classification: | Automated Sample Processing System |
Type: | Automated Sample Processing System |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Automated Sample Processing System
Application:
Sample processing for clinical diagnosis, epidemic surveillance, food safety, forensic identification, scientific research, etc., especially for samples of other virulent infectious diseases.
Features:
1. Safety: Automated Sample Processing system buitin UV lamp, and it can be used with a biologicalsafety cabinet, to efrectvely prevent aerosol pollution.
2. Efficient: Cooperative processing with dual robotic arms.
3. Convenient: Visual interface operation, easy to operate.
4. Compatibility: Compatible with a variety of pipette tips, deep well plates, sampling tubes.(including blood collection tubes) specifications.
5. Smart: One-key operation, smart dispensation.
Technical Parameters:
Model | BK-PR32 | BK-PR48 |
Throughput | 1~32 | 1~48 |
Processing Time | 32 samples/10min | 48 samples/16min |
Sample Type | Plasma, serum, whole blood, swab solution and other samples | |
Sample Rack | 1pc, 3*12 with locking device (compatible with a variety of sampling tubes) |
1pc, 6*8 with locking device(compatible with a variety of sampling tubes) |
Robot Arm | 1pc(Dispensation arm) | 2pcs(Dispensation arm and screw cap arm) |
Plate Position | 2pcs(Compatible with l-shaped and square boards) | 3pcs(Compatible with multi-specification deep-well plates) |
Tip Position | 3pcs(Including tip waste box position) | |
Reagent Rack | 1pc(4*2ml centrifuge tube+4*2ml freezing tube+4*5ml freezing tube) | |
Protective Function | Can be used in a biological safety cabinet External droplet catch tray design With air-tight and anti-drip design |
|
Liquid Detection | Pneumatic liquid level detection principle, intelligent detection of blocked needle | |
Pipetting Volume | 5~1000μl(1000/50μl tip) | |
Pipetting Accuracy | 10μl, CV≤1.5%, Accuracy≤6.0%, 50μl tip 50μl, CV≤1.0%, Accuracy≤2.0%, 1000μl tip 100μl, CV≤0.5%, Accuracy≤2.0%, 1000μl tip |
|
Power Supply | 220V, 50/60Hz; 110V, 60Hz | |
External Size(W*D*H)mm | 540*680*1113 | 827*794*1232 |
Net Weight | 56kg | 100kg |
Package Size(W*D*H)mm | 670*810*1314 | Main Instrument: 855*855*747 Base Cabinet: 960*925*975 |