Customization: | Available |
---|---|
Classification: | Biochemical Apparatus |
Type: | Biochemistry Analyzer |
Still deciding? Get samples of US$ 4610/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
The 400T/H Automatic Chemistry Analyzer is a state-of-the-art diagnostic tool designed for high-efficiency biochemical analysis.This advanced instrument boasts an impressive array of features tailored to meet the demands of modern laboratories.
Features:
1. 49 sample positions
2. 56 reagent positions.
3. 120 reaction cuvettes.
4. 400 tests per hour.
5. Probe with anti-collision function, liquid level dectection function.
Features:
1. Fully automated, Discrete, Random access,STAT priority,real-time dual wavelength.
2. Total open setting of items, supported different reagents from different countries.
3. Reversed optics and imported photo sensor make sure the precision of detection results.
4. Microsring injector imported from Germany assures the veracity of a trace of samples,
sample step by 0.1 UL, reagent step by0.5 UL
5. 24 hour non-stop working, convenient emergency function with inserting emergency at random.
6. Implementing analysis of ultra-trace reaction solution to 100 UL, and saving more reagents.
7. Discrete cuvette, preventing from cross-infection, and be washed and reused, making sure accurate result and save cost.
8. Perfect self-detection function makes sure the analyzer in order and accurate result.
9. Software is easy to learn and use. And test report can be printed.
10. Perfect function of quality control management is convenient for users to manage the clinical test result.
11. Detecting in the whole reaction process with reaction curve, convenient for users to monitor thoroughly
and scientifically analyze the detection data.
12. Supporting software can implement the function of detecting surplus reagent through liquid level
induction of reagent probe.
Model
|
BK-410
|
|
Overall Performance |
Throughput
|
400Tests/hour, ISE(Optional)
|
Analysis Method
|
End-point, Fixed-time, Rate(Kinetic), Turbidimetry
|
|
System
|
Optional open or close system, continuous working for 24 hours
|
|
Sample & Reagent System
|
Sample Positions
|
49sample positions
|
Reagent Position
|
56 reagent positions
|
|
Sample Volume
|
2-70μL
|
|
Reagent Volume
|
20-350μL
|
|
Probe
|
With anti-collision function, liquid level detection function
|
|
Probe Washing
|
Automatic washing interior and exterior
|
|
Reagent Cooling
|
2~12ºC for 24 hours
|
|
Reaction System
|
Temperature Control
|
Water bath37±0.2ºC, real-time monitoring
|
Cuvettes
|
120 reusable cuvettes
|
|
Mixer Probe
|
Independent stirring, automatic frequency conversion
|
|
Washing
|
Automatic cuvettes washing
|
|
Optical System
|
Light Source
|
Halogen lamp, water-cooling
|
Spectrophotometry
|
Post-spectral spectrophotometry
|
|
Wavelength
|
340~800nm, <±1nm
|
|
Absorbance
|
0~3.0Abs
|
|
Calibration & QC
|
Calibration
|
Linear: K factor, 1-point,2-point and multipoint linear Non-Linear: Spline, Polygon,Index,Ogarithm, Logit-4P, Logit-5P |
Quality Control
|
Real-time QC, Westgard multi rule, Cumulative Sum Check, Twin Plot(2D)
|
|
Data Management
|
Software
|
Windows 7/8/10, 32or 64bit
|
LIS System
|
Bi-direction, support HL7 protocol
|
|
Interface
|
LAN port access
|
|
Printer
|
External, multiple reporting mode available
|
|
Working Conditions
|
Power Supply
|
AC220V±10%, 50/60Hz(Standard), 110V±10% 60Hz(Optional), 650VA
|
Temperature
|
15~30ºC(±2ºC)
|
|
Water Consumption
|
Deionized water<10L/h
|
|
Humidity
|
40~85%
|
|
Size & Weight
|
External Size (W*D*H)
|
973*645*520mm
|
Net Weight
|
96kg
|
|
Package Size (W*D*H)
|
1130*735*1040mm
|
|
Gross Weight
|
141kg
|