Customization: | Available |
---|---|
After-sales Service: | Online |
Warranty: | 1 Year |
Still deciding? Get samples of US$ 4800/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
High Reliability Lab Gas Chromatograph
Features:
*7-inch LCD display indicates temperature, gas flow and heating curve.
*SelMest linction and automatic fault recognition function.
*8 pcs external extension interfaces can connect with control valve.
*Equipped with a flame ionization detector (FID).
*Multi-core and 32-bit embedded hardware system ensures the reliable operation.
*Synchronous external trigger function and the external signal (auto sampler, thermal analyzer, etc.) can start the equipment and workstation software at the same time.
*With chromatography workstation software to process data.
*Memory function: can store 20 sample test modes.
*One key start up function.
Model |
BK-GC112A | ||
Oven |
Capacity |
22L |
|
Temp. Range |
RT+5°C~400°C |
||
Temp. Accuracy |
±0.1 °C |
||
Temp. Program |
9 step |
||
Program Total Time |
9999.9 min |
||
Max Temp. Heating Rate |
60°C/min |
||
Sample Injector |
Temp. Range |
RT+7°C~420°C |
|
Column or split injector can be installed |
|||
Carrier gas flow control type: Constant pressure mode |
|||
Max. Quantity |
3pcs |
||
Pre Column Pressure Range |
0-400kpa |
||
Pre Column Pressure Accuracy |
0.1 kpa |
||
Flow Setting Range |
H2: 0-200ml/min |
||
N2: 0-150ml/min |
|||
Detector |
Temp. Range |
RT+7°C~420°C |
|
Hydrogen Flame Ionization Detector (FID) |
Logarithmic amplifier |
||
High voltage switch control |
|||
Baseline signal display |
|||
Detection limit: ^3*1012g/s (Sample: n-hexadecane) |
|||
Baseline noise: <5*1014A |
|||
Baseline draft:6*1 O ^A |
|||
Dynamic range: 107 |
|||
RSD:£3% |
|||
Thermal Conductivity Detector (TCD)(Optional) |
Sensitivity: 8000mv.ml/mg(Sample: n-hexadecane) |
||
Baseline noise: <0.05mV |
|||
Baseline draft: <0.15mV/30min |
|||
Dynamic range: 105 |
|||
Power Supply |
AC220V±10%, 50Hz |
||
Standard Accessory |
FID detector, Capillary injection system, Chromatography workstation software,RS232 port, LAN port |
||
Optional Accessory |
TCD detector, Chromatographic column, Gas purifier |
||
Package Size(W*D*H) |
735*725*835mm |
||
Gross Weight |
80kg |
Company Profile
Packaging & Shipping
Exhibition
Our Services
Contact us