Model
|
BK-UV1600G
|
BK-V1600G
|
Optical System
|
Single beam, grating 1200L/mm
|
|
Wavelength
|
190-1100nm
|
325-1100nm
|
Spectral Bandwidth
|
4nm
|
|
Wavelength Accuracy
|
±1.5nm
|
|
Wavelength Repeatability
|
≤0.2nm
|
|
Photometric Accuracy
|
+0.002Abs (0-0.5Abs), ±0.004A (0.5-1.0Abs), ±0.4%T (0-100%T)
|
|
Photometric Repeatability
|
0.001Abs(0-0.5Abs), 0.002Abs(0.5-1.0Abs), <0.2%T (0-100%T)
|
|
Stray Light
|
≤0.05%T(220nm,360nm)
|
|
Stability
|
±0.001A/h(500nm)
|
|
Photometric Range
|
-0.301-3.0A 0-200%T 0-9999C
|
|
Photometric Mode
|
A T C E
|
|
Wavelength Setting
|
Automatic
|
|
Output
|
RS232 printer data output, USB-B connection, wireless printing, mobile phone PC connection
|
|
Detector
|
Imported silicon photodiode
|
|
Standard Accessories
|
4*glass cuvette, 2*quartz cuvette
|
4*glass cuvette
|
Power Supply
|
100-220V 50/60Hz
|
|
External Size
|
466*365*200mm
|
|
Net Weight
|
11.25kg
|
|
Packing Size
|
590*480*315mm
|
|
Gross Weight
|
15.4kg
|
15kg
|
Worldwide Service:
1.Biobase India Branch Office.
2.Found in September, 2014, in Banglaore city, India.
3.BIOBASE warehouse in Jordan.
4.BIOBASE Business Incubator.
5.Found in 25th , July 2016 at Silicon Valley, California, USA.
6.BIOBASE Germany Branch Office (Under construction).
Pre-Sales Service:
1.Inquiry and consulting support.
2.Quick response with detailed quotation.
3.Confirm price, quantity, shipment, lead time and payment terms.
4.Make order and sealed PI.
5.Welcome to view our factory at any time.
6.OEM is welcomed.
After-Sales Service:
1.1-year warranty period.
2.Training how to install the machine, training how to use the machine.
3.Engineers available to service machinery overseas.