MicroPette Mechanical Pipette
All pipettes have been quality tested with calibration certificate. The quality control involves gravimetric testing of each pipette with distilled water at 22ºC.
Features for Micropette:
Single-channel Adjustable Volume Pipettes
|
||||||
Volume Range
|
Increment
|
Test Volume
(μl) |
Maximume Permissible Systematic error
(Inaccuracy) |
Maximume Permissible
Random error (Inaccuracy) |
||
|
|
|
%
|
μl
|
%
|
μl
|
0.1-2.5μl
|
0.05μl
|
2.5
|
2.50%
|
0.0625
|
2.00%
|
0.05
|
1.25
|
3.00%
|
0.0375
|
3.00%
|
0.0375
|
||
0.25
|
12.00%
|
0.03
|
6.00%
|
0.015
|
||
0.5-10μl
|
0.1μl
|
10
|
1.00%
|
0.1
|
0.80%
|
0.08
|
5
|
1.50%
|
0.075
|
1.50%
|
0.075
|
||
1
|
2.50%
|
0.025
|
1.50%
|
0.015
|
||
2-20μl
|
0.5μl
|
20
|
0.90%
|
0.18
|
0.40%
|
0.08
|
10
|
1.20%
|
0.12
|
1.00%
|
0.1
|
||
2
|
3.00%
|
0.06
|
2.00%
|
0.04
|
||
5-50μl
|
0.5μl
|
50
|
0.60%
|
0.3
|
0.30%
|
0.15
|
25
|
0.90%
|
0.225
|
0.60%
|
0.15
|
||
5
|
2.00%
|
0.1
|
2.00%
|
0.1
|
||
10-100μl
|
1μl
|
100
|
0.80%
|
0.8
|
0.15%
|
0.15
|
50
|
1.00%
|
0.5
|
0.40%
|
0.2
|
||
10
|
3.00%
|
0.3
|
1.50%
|
0.15
|
||
20-200μl
|
1μl
|
200
|
0.60%
|
1.2
|
0.15%
|
0.3
|
100
|
0.80%
|
0.8
|
0.30%
|
0.3
|
||
20
|
3.00%
|
0.6
|
1.00%
|
0.2
|
||
50-200μl
|
1μl
|
200
|
0.60%
|
1.2
|
0.15%
|
0.3
|
100
|
0.80%
|
0.8
|
0.30%
|
0.3
|
||
50
|
1.00%
|
0.5
|
0.40%
|
0.2
|
||
100-1000μl
|
5μl
|
1000
|
0.60%
|
6
|
0.20%
|
2
|
500
|
0.70%
|
3.5
|
0.25%
|
1.25
|
||
100
|
2.00%
|
2
|
0.70%
|
0.7
|
||
200-1000μl
|
5μl
|
1000
|
0.60%
|
6
|
0.20%
|
2
|
500
|
0.70%
|
3.5
|
0.25%
|
1.25
|
||
200
|
0.90%
|
1.8
|
0.30%
|
0.6
|
||
1000-5000μl
|
50μl
|
5000
|
0.50%
|
25
|
0.15%
|
7.5
|
2500
|
0.60%
|
15
|
0.30%
|
7.5
|
||
1000
|
0.70%
|
7
|
0.30%
|
3
|
||
2-10ml
|
0.1ml
|
10ml
|
0.60%
|
60
|
0.20%
|
20
|
5ml
|
1.20%
|
60
|
0.30%
|
15
|
||
2ml
|
3.00%
|
60
|
0.60%
|
12
|
8-channel Adjustable Volume Pipettes
|
||||||||
Volume Range
|
Increment
|
Test Volume
(μl) |
Maximume Permissible Systematic error
(Inaccuracy) |
Maximume Permissible Random error
(Inaccuracy) |
||||
|
|
|
%
|
μl
|
%
|
μl
|
||
0.5-10μl
|
0.1μl
|
10
|
1.50%
|
0.15
|
1.50%
|
0.15
|
||
5
|
2.50%
|
0.125
|
2.50%
|
0.125
|
||||
1
|
4.00%
|
0.04
|
4.00%
|
0.04
|
||||
5-50μl
|
0.5μl
|
50
|
1.00%
|
0.5
|
0.50%
|
0.25
|
||
25
|
1.50%
|
0.375
|
1.00%
|
0.25
|
||||
5
|
3.00%
|
0.15
|
2.00%
|
0.1
|
||||
50-300μl
|
5μl
|
300
|
0.70%
|
2.1
|
0.25%
|
0.75
|
||
150
|
1.00%
|
1.5
|
0.50%
|
0.75
|
||||
50
|
1.50%
|
0.75
|
0.80%
|
0.4
|
12-channel Adjustable Volume Pipettes
|
||||||||
Volume Range
|
Increment
|
Test Volume
(μl) |
Maximume Permissible Systematic error
(Inaccuracy) |
Maximume Permissible Random error
(Inaccuracy) |
||||
|
|
|
%
|
μl
|
%
|
μl
|
||
0.5-10μl
|
0.1μl
|
10
|
1.50%
|
0.15
|
1.50%
|
0.15
|
||
5
|
2.50%
|
0.125
|
2.50%
|
0.125
|
||||
1
|
4.00%
|
0.04
|
4.00%
|
0.04
|
||||
5-50μl
|
0.5μl
|
50
|
1.00%
|
0.5
|
0.50%
|
0.25
|
||
25
|
1.50%
|
0.375
|
1.00%
|
0.25
|
||||
5
|
3.00%
|
0.15
|
2.00%
|
0.1
|
||||
50-300μl
|
5μl
|
300
|
0.70%
|
2.1
|
0.25%
|
0.75
|
||
150
|
1.00%
|
1.5
|
0.50%
|
0.75
|
||||
50
|
1.50%
|
0.75
|
0.80%
|
0.4
|
Fixed Volume Pipettes
|
||||||||
Volume Range
|
Increment
|
Test Volume
(μl) |
Maximume Permissible Systematic error
(Inaccuracy) |
Maximume Permissible Random error
(Inaccuracy) |
||||
|
|
|
%
|
μl
|
%
|
μl
|
||
5μl
|
-
|
5μl
|
1.3%
|
0.065
|
1.2%
|
0.06
|
||
10μl
|
-
|
10μl
|
0.8%
|
0.08
|
0.8%
|
0.08
|
||
20μl
|
-
|
20μl
|
0.6%
|
0.12
|
0.5%
|
0.1
|
||
25μl
|
-
|
25μl
|
0.5%
|
0.125
|
0.3%
|
0.075
|
||
50μl
|
-
|
50μl
|
0.5%
|
0.25
|
0.3%
|
0.15
|
||
100μl
|
-
|
100μl
|
0.5%
|
0.5
|
0.3%
|
0.3
|
||
200μl
|
-
|
200μl
|
0.4%
|
0.8
|
0.2%
|
0.4
|
||
250μl
|
--
|
250μl
|
0.4%
|
1.0
|
0.2%
|
0.5
|
||
500μl
|
-
|
500μl
|
0.3%
|
1.5
|
0.2%
|
1.0
|
||
1000μl
|
-
|
1000μl
|
0.3%
|
3.0
|
0.2%
|
2.0
|
||
2000μl
|
-
|
2000μl
|
0.3%
|
6.0
|
0.15%
|
3.0
|
||
5000μl
|
-
|
5000μl
|
0.3%
|
15
|
0.15%
|
7.5
|
Company information:
Founded in 1999, having over 5000 employees , BIOBASE Group is a new high-tech enterprise focusing on development, production and management of scientific equipment, In Vitro Diagnosis (IVD) instrument and reagent.
BIOBASE Group is specialized in products of 8 areas including medical diagnosis, biosafety protection, disinfection and sterilization, water purification system, infant care products, cold chain products, software products, clean room project.
Contact: