Customization: | Available |
---|---|
Customized: | Customized |
Certification: | / |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Infant Radiant Warmer is one of the important equipment used in medical institutions. It is mainly used for constant temperature cultivation, temperature resuscitation, oxygen delivery, rescue, hospitalization and observation of sick infants and premature infants. It is used in hospitals, confinement centers, neonatology, etc.
Features:
LCD display
Double control modes of bed temperature and skin temperature
Double-layer thermostatic cover, automatic air curtain device
Humidity control function
Air & Oxygen blender
Oxygen supply system
Baby cot is infinitely adjustable
24 kinds of sound and light alarm, comprehensive protection of baby safety
General parameters |
Item |
Specifications |
Power supply |
AC220V,50/60Hz(Standard) 1320VA; 110V 60Hz(Optional) |
|
Size(L*W*H) |
950*645*(1250-11450) mm |
|
Maximum heater output power(220VAC) |
Air chamber heater:330W; Water tank heater:155W |
|
Auxiliary network power output |
AC220V/50Hz, both maximum permissible currents are 220V 1.5A |
|
Heater indication |
0~100%, 10% increasing amount |
|
Temperature control mode |
Bed temperature mode; skin temperature mode |
|
Bed temperature control range |
25.0~37.0ºC; 37.1~39.0ºC(Span mode) |
|
Skin temperature control range |
34.0~37.0ºC;37.1~38.0ºC(Span mode) |
|
Skin temperature sensor display range |
5.0~65.0ºC |
|
Heating time |
≤ 50min |
|
Temperature variability |
≤ 0.5ºC |
|
Temperature uniformity (mattress in horizontal position) |
≤ 0.8ºC |
|
Temperature uniformity (mattress in tilted position) |
≤1.0ºC |
|
Difference between average incubator temperature and control temperature |
≤1.5ºC |
|
Skin temperature sensor accuracy |
±0.3ºC |
|
Humidity control |
The running time of the tank after it is filled with distilled water |
Humidity below 70%RH, at least 24h |
Water tank capacity |
1500ml |
|
Humidity display range |
0%RH~99%RH |
|
Humidity control range |
CAH%RH~90%RH(1% per adjustment) |
|
Humidity control accuracy |
±10% |
|
Oxygen concentration control |
Oxygen concentration display range |
0%~99% |
Oxygen concentration display resolution |
1% |
|
Oxygen concentration control accuracy |
±5% oxygen volume concentration |
|
Oxygen concentration setting range |
21% ~ 60% (1% per adjustment) |
|
Oxygen sensor service life |
The maximum use time is 10000h at 100% oxygen concentration |
|
Transport storage |
Temperature |
-20ºC~+55ºC |
Atmospheric pressure range |
500hPa~1060hPa |
|
Relative humidity |
≤93%RH |
|
Packing size/ Gross Weight |
Base packing size: 1200*800*870mm 50Kg Body packing size: 1100*710*855mm 60Kg |
|
The work environment |
Relative humidity |
30%RH~75%RH |
Environment temperature |
+20ºC~+30ºC |
|
Pressure range |
700hPa~1060hPa |
|
Service altitude |
≤2000m |
|
Overvoltage category |
|
|
Pollution level |
2 |
|
Ambient air velocity |
≤ 0.3 m/s |
|
Other indicators |
Ambient noise in the infant compartment of the incubator |
No more than 55dB(A) at stable temperature |
Carbon dioxide concentration in the infant compartment |
A mixture of 4% carbon dioxide and air was fed into an 8mm diameter tube in the vertical direction from the mat to the top of the box at 750mL /min to a point 10cm above the mattress center and 10cm above the center. The concentration was less than 0 5% after stabilization |
|
The velocity of air flow over the baby mattress |
≤0.35m/s |
|
General alarm |
Power failure alarm, fan alarm, sensor alarm, deviation alarm, over temperature alarm, water tank placement error alarm, water shortage alarm |
|
System alarm |
Mcu1 and Mcu2 communication error alarm, Mcu2 and Mcu3 communication error alarm, analog-to-digital conversion chip 2544 failure alarm, analog-to-digital conversion chip 2548 failure alarm, over temperature alarm, temperature heating system failure, humidity heating system failure |
Configuration:
Model |
BK-3201(I) |
BK-3201(II) |
Temp. Control |
Bed/Skin temperature |
Bed/Skin temperature |
Display |
LCD |
LCD |
Bed Temp. Control Range |
25ºC-39ºC |
25ºC-39ºC |
Skin Temp. Control Range |
34ºC-38ºC |
/ |
Humidity Control Range |
CAH%~99%RH |
CAH%~99%RH |
Oxygen Control Range |
21%-99% |
/ |
Cabinet |
Yes |
Yes |