Introducton:
Real-time PCR is used for sensitive, specific detection and quantification of nucleic acid targets. We have developed powerful assay design algorithms, optimized qPCR regent, intuitive data analysis software, and flexible instrumentation to help harness the power of qPCR across a rich and diverse set of applications. Explore our robust solutions for your qPCR-based research.
Application:
It can be widely used for Infectious disease research, Food pathogen detection, Waterborne pathogen detection, Stem cell research, Oncology and genetic disease research, Plant sciences and agricultural biotechnology.
Working Principle:
The temperature step change is controlled by the semiconductor peltier to realize PCR amplification. Use high-sensitivity PD unit to detect fluorescence. Program control channel switching, non-contact excitation/detection on the top structure, coordinated with motor control X and Y axis movement to achieve 96-hole scanning. Finally, accurate analysis is carried out through powerful software.
Features:
1. Excellent temperature control performance of the instrument, Module Max heating rateheating rate 7.0ºC/s.
2. No edge effect, no optical path correction, top excitation/detection, non-contact measurement.
3. Fluorescence detection adopts PD sensor with high sensitivity.
4. Long-life LED light source, stable emission wavelength, maintenance-free.
5. 4/6-channel fluorescence detection, no cross interference between channels.
6. User-friendly and fully functional software, flexible program setting, comprehensive analysis and reporting functions, all the parameters can be stored.
Model |
LEIA-X4 |
|||
Sample Capacity |
96*0.1ml PCR plate, 12*8-strip tubes, 96*0.1ml single tube(Transparent cover) |
|||
Reaction System |
10~50μl |
|||
Dynamics Range |
1~1010 copies |
|||
Channel |
4 |
|||
Emission Light |
LED |
|||
Detector |
MPPC |
|||
Detection Path |
F1 |
F2 |
F3 |
F4 |
Suitable Probe/Dye |
FAM/SYBR GREEN
|
VIC/JOE/ HEX/TET |
ROX/TEXAS-RED |
Cy5
|
Excitation Wavelength |
455~680nm |
|||
Detection Wavelength |
510~730nm |
|||
Fluorescence Detection Repeatability |
CV≤2% |
|||
Fluorescence Detection Accuracy |
CV≤3% |
|||
Fluorescence Detection Linearity |
r≥0.995 |
|||
Module Temp. Range |
4~99ºC(Resolution: 0.1ºC) |
|||
Ramp Rate |
5.0ºC/s(Max) |
|||
Temp. Accuracy |
±0.3ºC |
|||
Temp. Uniformity |
≤±0.3ºC |
|||
Temp. Control Mode |
Block mode |
|||
Gradient Temp. Range | 1~36ºC | |||
Hot-lid Temp. Range |
100ºC, automatic hot-lid |
|||
Scanning Mode |
Full plate scanning |
|||
Programming |
Max 100 segments for each program, max 99 cycles |
|||
Operation Mode |
Continuous |
|||
Scanning Time |
8.5s |
|||
Special Function |
Absolute quantitative automatic analysis, relative quantification, SNP analysis, melting curve analysis, etc |
|||
Operation System |
Microsoft: Windows10 |
|||
Power Supply |
220V, 50/60Hz; 110V, 60Hz |
|||
Dimension(L*W*H)mm |
375*490*365 |
|||
Port Method |
USB Port |
|||
Packing Size(L*W*H)mm |
645*565*605 |
|||
Gross Weight(kg) |
45 |