Customization: | Available |
---|---|
Classification: | Biochemical Apparatus |
Type: | Mobile PCR Laboratory |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Area arrangement
1.The PCR laboratory was designed in strict accordance with the three functional divisions: reagent preparation area,specimen preparation area and product amplification and analysis area, with independent buffer room.In addition, configure the washroom and equipment room.
PCR Laboratory Equipment
Capsule | 1.Single box structure, factory modular form a complete set. 2.Dimensions: 12192mm×2438mm×2896mm (l/D), standard container body, easy to transport. 3.The wall is made of 2mm-thick reinforced wavy steel plate, and the floor is made of 5cm-thick special wood plate with high strength, which is suitable for transportation under various terrain conditions. 4.6mm thick high-quality steel support frame, stable and durable; 5.Stainless steel outer door, double door design, with access control system; 6.The surface is treated with rust removal, anti-rust paint and container special paint for three times. It is all-weather and durable. |
Inside cabin | 1.The ceiling and internal partition walls are made of 50mm laboratory dedicated antibacterial purification board with a thickness of up to 0.426mm, which has good antibacterial effect, fire prevention, moisture-proof. 2.The Angle of Yin and Yang in the box is equipped with a special circular aluminum profile for the clean room, which ensures that the inside of the box is smooth and has no dead Angle, no ash accumulation, and is not easy to produce mold. 3.Special steel clean sealed door, with glass observation window, two doors in buffer room have interlock device, and equipped with emergency opening device, can be manually opened in emergency situations such as power failure; 4.The ground is made of 2mm PVC floor leather, which is fireproof, antibacterial, wear-resistant and non-slip. |
Area arrangement | 1.The PCR laboratory was designed in strict accordance with the three functional divisions: reagent preparation area, specimen preparation area and product amplification and analysis area, with independent buffer room.In addition, configure the washroom and equipment room. 2.The built-in PCR corridor facilitates the observation of the operation of each partition 3.The pressure gradient decreases step by step according to the direction of reagent preparation area, specimen preparation area and product amplification analysis area, and the airflow trend is one-way flow. 4.Each area is equipped with transfer window for goods transfer. The transfer window has the function of interlock and ultraviolet lamp disinfection. 5.Each area is equipped with LED clean lamps, which are energy saving and easy to maintain, and can effectively prevent the breeding of bacteria. The room illumination meets the standard requirements. 6.It is equipped with special ultraviolet lamp and time-delay function to protect the safety of experimental personnel. 7.A pressure differential meter is installed at the door of the buffer room to display the dynamic pressure in the room in real time. |
Ventilation | 1.Strictly follow the PCR laboratory air supply and exhaust design to ensure the air volume and ventilation frequency in the experimental area; 2. The combined clean air conditioning unit adopts primary and medium efficiency two-stage filter air supply for the fresh air unit and high-efficiency filter for the terminal air supply port to carry out three-stage filter air supply to ensure the cleanliness of the room meets the requirements. 3. The combined air conditioning unit carries out initial and medium efficiency, surface cooling and reheating system, filtration and temperature regulation for fresh air, which can meet the whole weather conditions of -20 ° C to 40 ° C and ensure the indoor temperature is constant. 4. The internal temperature of the laboratory is 18ºC ~ 26ºC, and the relative humidity does not exceed 70%. 5. Independent exhaust system, laboratory negative pressure high efficiency exhaust fan equipped with high efficiency exhaust filter, can effectively filter indoor biological aerosol, to avoid environmental pollution. |
Water supply and drainage system |
1.The clean water tank and waste water treatment equipment are integrated between the equipment to maintain the normal operation of the laboratory under the condition of water cut off. 2.Equipped with 260L capacity of clean water tank; 3.The drainage of the laboratory is collected centrally and discharged after inactivated treatment. The discharge standard meets relevant national standards. 4.The laboratory wastewater treatment equipment adopts neutralization and precipitation, compound disinfection, UV ultraviolet sterilization, ozone oxidation, multi-medium filtration and other technologies to treat all kinds of pollutants in wastewater; |
Laboratory table | 1. The mesa is 12.7mm thick and the edges of the high-quality solid core physicochemical board are thickened to 25.4mm, which is resistant to corrosion, abrasion and strong acid and alkali. 2. The frame is made of 1.2mm high-quality cold-rolled steel plate, the surface is sprayed with epoxy resin electrostatic, the surface is smooth and smooth, without falling off bubble pits, indentation and scratch pitting points, cracks and cutting edges and other defects;The bottom is supported by adjustable feet, used for horizontal trimming joints of mesa. 3.The cabinet is made of 1.2mm thick steel plate. The surface is defatted, washed, pickled, neutralized, phosphated and washed by high pressure washing and drying. The surface is cured by epoxy resin spraying powder at high temperature. 4. The accessories are of the exclusive brand of the high quality laboratory, equipped with the special faucet for the laboratory, PP sink, eye washer, etc. |
Controlling system | 1.Visual intelligent control system, with room temperature and humidity, pressure gradient, lighting, uv lamp control; 2. With dynamic pressure control ability, it can automatically balance the change of pressure difference caused by air conditioning filter screen and high efficiency filter blockage, and display the running state of air supply and exhaust unit; 3. It has the function of exhaust air efficient clogging detection, realizing the alarm function when clogging occurs and reminding to replace. 4.With pressure gradient monitoring function, automatic alarm to remind users when the data is out of range;The room will monitor the temperature and humidity data in real time, and automatically alarm the user when the data is out of range; 5.With the function of pre-sterilization and intelligent sterilization, the abnormal situation always automatically alarm to remind users. |